Assignement: Whole Brain Module
Case # 2
Patient Name: John Whole Brain
Treatment Site: SSD Whole Brain
Field Size: X=26 Y=18
Collimator Rotation 30 degrees
Calculation Point: The Isocenter
RX: 3000 cGy in 10 Fractions; 6MV
Beam arrangement: 2 Lateral Beams
IFD: 16 CM
The IFD (Inner Field Distance) is the thickness of the patient through the planning area.
If the thickness is 16 cm, and the patient is setup midplane, then we know that ½ of the IFD is the distance from the surface of the skin to the CA; in this case, 8cm. Since this case is set-up SSD, the SSD will be 100 and 8cm will be our thickness.
 |
Page 10 of 11 |
 |